中文摘要: 建立了氟化氢铵消解样品,电感耦合等离子体质谱法测定地球化学样品中锂、铷和铯的方法。方法经国家土壤和水系沉积物标准参考物质验证,方法的检出限锂为0.012μg/g、铷为0.038μg/g、铯为0.045μg/g;准确度(相对误差)锂为2.51%~4.77%、铷为2.84%~3.41%、铯为4.00%~7.77%;精密度锂为0.41%~1.28%、铷为0.32%~0.94%、铯为1.28%~2.54%;能够满足地球化学样品分析测试的质量要求。并且有效的避免了铷和铯在样品消解过程中的挥发损失。方法具有操作简便、测试成本低、分析效率高等特点,适合于批量样品的分析测试。
中文关键词: 氟化氢铵,消解,电感耦合等离子体质谱仪
Abstract:A method for the determination of lithium, rubidium and cesium in Geochemical Samples by inductively coupled plasma mass spectrometry (ICP-MS) after digestion with ammonium hydrogen fluoride was established. The method was verified by the national standard reference material of soil and stream sediment. The detection limits of lithium, rubidium and cesium were 0.012 μ g / g, 0.038 μ g / g and 0.045 μ g / g, respectively; The accuracy (relative error) of lithium is 2.51% - 4.77%, rubidium is 2.84% - 3.41%, cesium is 4.00% - 7.77%;The precision of lithium is 0.41% - 1.28%, rubidium is 0.32% - 0.94%, cesium is 1.28% - 2.54%.The quality requirements of geochemical sample analysis and testing can be met. Moreover, the volatilization loss of rubidium and cesium in the sample digestion process is effectively avoided. The method has the advantages of simple operation, low test cost and high analysis efficiency, which is suitable for the analysis and test of batch samples.
keywords: Ammonium hydrogen fluoride digestion inductively coupled plasma mass spectrometry lithium rubidium cesium
文章编号: 中图分类号: 文献标志码:
基金项目:
作者 | 单位 | |
任梦阳* | 长春科技学院 | ren203266607@163.com |
Author Name | Affiliation | |
renmengyang | Changchun University of science and technology | ren203266607@163.com |
引用文本:
任梦阳.氟化氢铵消解-电感耦合等离子体质谱(ICP-MS)法测定地球化学样品中的锂铷铯[J].中国无机分析化学,2021,11(4):54-57.
renmengyang.REN Mengyang[J].Chinese Journal of Inorganic Analytical Chemistry,2021,11(4):54-57.
任梦阳.氟化氢铵消解-电感耦合等离子体质谱(ICP-MS)法测定地球化学样品中的锂铷铯[J].中国无机分析化学,2021,11(4):54-57.
renmengyang.REN Mengyang[J].Chinese Journal of Inorganic Analytical Chemistry,2021,11(4):54-57.