投稿时间:2024-11-27 修订日期:2024-12-25
中文摘要: 配备动态反应池(DRC)的三重串联四极杆电感耦合等离子体质谱仪成功应用于准确测定金属铜/镍/钴样品中的磷。该方法利用31P+与池中O2的反应,通过在m/z 47处检测31P+的氧化物离子31P16O+从而实现磷的定量,消除了多原子离子干扰,进一步优化了DRC系统工作条件的影响,在O2流速为0.9mL/min的条件下获得31P16O+的最佳信噪比。采用标准加入法的测定方式克服金属铜/镍/钴基体产生的基体效应,实验结果与国家标准方法分光光度法比对结果基本一致。方法线性范围为0.0001%~0.0050%,线性相关系数均在0.9995以上,金属铜/镍/钴中磷的检出限均在0.3ng/mL以下。实际样品测定结果的相对标准偏差在4.14%~13.84%,回收率在94.9%~118.5%。
中文关键词: 三重串联四极杆电感耦合等离子体质谱仪 金属铜/镍/钴 DRC O2
Abstract:A triple tandems quadrupole inductively coupled plasma mass spectrometer equipped with a dynamic reaction cell (DRC) has been successfully applied to accurately determine phosphorus in copper/nickel/cobalt metal samples. This method utilizes the reaction between 31P+and O2 in the pool to quantify phosphorus by detecting the oxide ion 31P16O+of 31P+at m/z 47, eliminating the interference of multiple atomic ions and further optimizing the working conditions of the DRC system. The optimal signal-to-noise ratio of 31P16O+ is obtained under the condition of O2 flow rate of 0.9 mL/min. The standard addition method was used to overcome the matrix effect caused by the metal copper/nickel/cobalt matrix, and the experimental results were basically consistent with the results obtained by the national standard method spectrophotometry. The linear range of the method is 0.0001%~0.0050%, and the linear correlation coefficients are all above 0.9995. The detection limit of phosphorus in copper/nickel/cobalt metals is below 0.3ng/mL. The relative standard deviation of the actual sample measurement results is between 4.14% and 13.84%, and the recovery rate is between 94.9% and 118.5%.
keywords: Triple Tandem Quadrupole Inductively Coupled Plasma Mass Spectrometry Metal Copper/Nickel/Cobalt DRC O2
文章编号: 中图分类号: 文献标志码:
基金项目:矿冶科技集团有限公司青年科技创新基金项目资助(课题编号04-2403);国家重点研发计划项目资助(2021YFC2903101)
引用文本:
孙梦荷,刘春峰,方迪,苏春风,董梦君,陈殿耿.三重串联四极杆 - 电感耦合等离子体质谱( ICP-MS)法测定金属铜 / 镍 / 钴中的磷[J].中国无机分析化学,2025,15(3):406-413.
sunmenghe,LIU Chunfeng,FANG Di,SU Chunfeng,DONG Mengjun,CHEN Diangeng.Determination of Phosphorus in Copper/Nickel/Cobalt Metal by Triple Tandem Quadrupole-Inductively Coupled Plasma Mass Spectrometry[J].Chinese Journal of Inorganic Analytical Chemistry,2025,15(3):406-413.
孙梦荷,刘春峰,方迪,苏春风,董梦君,陈殿耿.三重串联四极杆 - 电感耦合等离子体质谱( ICP-MS)法测定金属铜 / 镍 / 钴中的磷[J].中国无机分析化学,2025,15(3):406-413.