Simultaneous Determination of Cd, Co, Cu, Mg, Mn, Ni, Pb, Zn And Ca in Laterite Nickel Ore by Inductively Coupled Plasma Atomic Emission Spectrometry (ICP-AES)
Received:December 27, 2010  Revised:March 17, 2011
View Full Text  View/Add Comment  Download reader
DOI:10.3969/j.issn.2095-1035.2011.02.0006
KeyWord:laterite nickel ore;ICP-AES; multi-elements; simultaneous determination
        
AuthorInstitution
HE Feiding 桂林理工大学,广西桂林
LI Huachang 北京矿冶研究总院,北京
FENG Xianjin 北京矿冶研究总院,北京
Hits: 3266
Download times: 2142
Abstract:
      A new analytical method was proposed to simultaneously determine Cd, Co, Cu, Mg, Mn, Ni, Pb, Zn and Ca in laterite nickel ore by inductively coupled plasma atomic emission spectrometry (ICP-AES). The samples were first treated by hydrochloric acid, nitric acid and hydrofluoric acid and then determined by ICP-AES under optimal instrument conditions with hydrochloric acid as the testing medium. The detection limits for each element were 0~0.0033 μg/mL and the relative standard deviations (RSD,n=6)were 0.15%~6.95%. The recoveries for testing the spiked samples were between 91.4%~107.2%. The comparison test showed that the values determined by ICP-AES were consistent with those by flame atomic absorption spectrometry.
Close