Determination of Primary Elements in Potassium and Sodium Feldspar by X-ray Fluorescence Spectrum Method (XRF)
Received:November 29, 2011  Revised:March 27, 2012
View Full Text  View/Add Comment  Download reader
DOI:doi:10.3969/j.issn.2095-1035.2012.02.0008
KeyWord:XRF; fusion sample preparation; feldspar
           
AuthorInstitution
SHANG Ying 河北钢铁集团邯钢有限公司 技术中心,河北 邯郸
WANG Binguo 河北钢铁集团邯钢有限公司 技术中心,河北 邯郸
ZHAO Jing 河北钢铁集团邯钢有限公司 技术中心,河北 邯郸
XU Jing 河北钢铁集团邯钢有限公司 技术中心,河北 邯郸
Hits: 3471
Download times: 95
Abstract:
      In this paper, the contents of Na2O, Al2O3, SiO2, K2O, CaO, Fe2O3 andTiO2 in potassium and sodium feldspar were determined by XRF. Samples were prepared by fusion method. A series of standard solutions for calibration were prepared mainly with the national standards of potassium and sodium feldspar and supplementary with clay, slag and silica standards. The results show that standard deviation of this method is 0.065%~2.9%, which is basically consistent with the standard value. This method, due to its better accuracy, can be used as a routine method for analyzing feldspar-like aluminosilicate minerals and replace the wet chemical analysis method.
Close