Microwave digestion and determination of manganese,phosphorus,nickel,copper,chromium and vanadium in high-speed tool steel by ICP-AES |
Received:January 22, 2014 Revised:February 12, 2014 |
View Full Text View/Add Comment Download reader |
DOI: |
KeyWord:Microwave digestion; ICP-AES; high-speed tool steel; matrix matching |
Author | Institution |
HE Xiao-hu |
广西冶金产品质量监督检验站 |
GAN Ri-xing |
广西冶金产品质量监督检验站 |
WEI Li |
广西冶金产品质量监督检验站 |
|
Hits: 1067 |
Download times: 209 |
Abstract: |
High speed tool steel is a kind of high alloy tool steel with high carbon, it is difficult to be dissolved at room temperature. In this paper, by using microwave digestion improves digestion temperature and pressure, samples were full digested in hydrofluoric acid, sulfuric acid and aqua regia medium, by adding saturated boric acid solution complexation excess of hydrogen fluoride acid, and matrix matching to eliminate the influence of the iron matrix by ICP-AES method for simultaneous determination of manganese, phosphorus, copper, nickel,,chromium, vanadium content.. Determination of high speed tool steel standard, the measured value is consistent with the certified values. The precision of this method is in 0.55% ~ 4.06%,the recoveries were 95.6% ~ 114.8%. This method can meet the requirements of the measurement. |