Determination of Total Chromium in Industrial Waste Water by Atomic Absorption Spectrometry and Atomic Emission Spectrometry |
Received:February 14, 2014 Revised:April 04, 2014 |
View Full Text View/Add Comment Download reader |
DOI:doi:10.3969/j.issn.2095-1035.2014.02.001 |
KeyWord:atomic absorption spectrometry; atomic emission spectrometry; total chromium; comparison |
Author | Institution |
XU Guojin |
宁波市环境监测中心,浙江 宁波 |
FAN Yingguo |
宁波市环境监测中心,浙江 宁波 |
ZHAO Qian |
宁波市环境监测中心,浙江 宁波 |
|
Hits: 2205 |
Download times: 0 |
Abstract: |
In order to investigate the similarities and differences between atomic absorption spectrometry and atomic emission spectrometry for the determination of total chromium in industrial waste water, we separately applied these two methods to determine the total chromium in industrial waste water and made a comparison of the testing results. Sample pre-treatment, the standard curves, the detection limits of the method, accuracy, precision, interference and its elimination were compared between the two methods. Furthermore, the comparison results were analyzed using t-test, showing the two methods were in good agreement. However, the performance indexes of atomic emission spectrometry were superior in all aspects to atomic absorption spectrometry. |