Determination of Trace Thallium in Zn-Matrix Material by Graphite Furnace Atomic Absorption Spectrometry |
Received:November 11, 2013 Revised:December 10, 2013 |
View Full Text View/Add Comment Download reader |
DOI:doi:10.3969/j.issn.2095-1035.2014.02.002 |
KeyWord:thallium; graphite furnace atomic absorption spectrometry; zinc matrix material; matrix modifier |
Author | Institution |
CHENG Jian |
深圳中金岭南公司 丹霞冶炼厂,广东 仁化 |
|
Hits: 1798 |
Download times: 0 |
Abstract: |
The sample was dissolved by B -HNO3-A and palladium nitrate was used as matrix modifier. The optimal conditions of instrumental analysis were tested and the results as follows: ashing temperatures were 500 ℃ and atomization temperature was 1 800 ℃. Under the optimal conditions, graphite furnace AAS method was successfully applied to determine the trace thallium in zinc concentrates that was used as zinc-matrix material for refining zinc by hydrometallurgical method , calcine, zinc oxide, zinc powder, etc. The results were basically consistent with those by ICP-MS method. The relative standard deviations (RSD, n ≤ 8) were 1.7%~7.8% and the recovery rates were 99%~103%. |
Close |