Determination of Major and Minor Components in Magnesia by X-ray Fluorescence Spectrometry
Received:June 30, 2014  Revised:August 19, 2014
View Full Text  View/Add Comment  Download reader
DOI:doi:10.3969/j.issn.2095-1035.2014.04.011
KeyWord:X-ray fluorescence spectrometry; magnesia; fusion sampling
     
AuthorInstitution
WU Zengsheng 中国第一重型机械股份公司 天津重型装备工程研究有限公司,天津
LIU Zhimin 中国第一重型机械股份公司 天津重型装备工程研究有限公司,天津
Hits: 1288
Download times: 0
Abstract:
      With certified reference materials of magnesia as calibration samples,a new method was established for the determination of major and minor components including MgO, Al2O3, SiO2, CaO, P2O5 and Fe2O3 in magnesia by X-ray fluorescence spectrometry. The preparation of testing sample and calibration sample were carried out by fusion method. The optimized experimental conditions were obtained as follows: a mixed solution of Li2B4O7-LiBO2(67+33) was used as fusing agent, 1 mL LiBr was used as remover, the fusion temperature was 1 100 ℃, and the fusion time was 20 min. The relative standard deviation (RSD) was less than 3%.The determination results were consistent with the certified values.