Flame atomic absorption spectrometry determination of the amount oflead in copper the original tailings |
Received:January 19, 2015 Revised:January 19, 2015 |
View Full Text View/Add Comment Download reader |
DOI: |
KeyWord:atomic absorption spectrometry; Copper original tailings; lead |
Author | Institution |
Zhang guanghua |
北方铜业股份有限公司计量检验部 山西垣曲 |
Ma lijun |
北方铜业股份有限公司计量检验部 山西垣曲 |
|
Hits: 1232 |
Download times: 220 |
Abstract: |
: copper tailings of silicon content is higher, according to this feature, this paper puts forward using saturated ammonium hydrogen fluoride, hydrochloric acid, nitric acid to dissolve sample, in 5% nitric acid medium, on the atomic absorption spectrometer, the use of air - acetylene flame to determine the lead, copper is established on the determination of lead in the original tailings method. The results show that the mass concentration of lead within 0.05 ~ 3 mu g/mL and absorbance A good linear relationship, the relative standard deviation of 0.92% to 2.39%, standard addition recovery rate from 96.0% to 106.0%, three times the blank solution method to 11 times the standard deviation of the determination of value as A detection limit, detection limit of 0.08 u g/mL. |