Determination of Beryllium in Beryllium Associated Uranium Ore by Inductively Coupled Plasma Optical Emission Spectrometry |
Received:July 11, 2014 Revised:March 17, 2015 |
View Full Text View/Add Comment Download reader |
DOI:doi:10.3969/j.issn.2095-1035.2015.02.006 |
KeyWord:beryllium;ICP-OES; standard addition; acid soluble; alkali melting; beryllium associated uranium ore |
Author | Institution |
DU Guirong |
核工业北京化工冶金研究院,北京 |
ZHAO Jie |
西安中核蓝田铀业公司,西安 |
DING Hongfang |
核工业北京化工冶金研究院,北京 |
LIU Yang |
核工业北京化工冶金研究院,北京 |
WANG Chunye |
核工业北京化工冶金研究院,北京 |
|
Hits: 2324 |
Download times: 0 |
Abstract: |
In this study, an inductively coupled plasma optical emission spectrometric method for determining beryllium in beryllium associated uranium ore was established. Matrix interference from potassium pyrosulfate, which was added to the digestion solution with alkali fusion and acid dissolution, was investigated, and measures to eliminate the matrix interference were discussed. The results showed that sample weight, potassium pyrosulfate adding amount, constant volume and dilution ratio should be control to effectively eliminate the matrix interference. The relative standard deviation was 3.4%(n=6) . The recovery was 98.7%~104.4%. The proposed method is reliable, accurate, and has been proved by practical application. |
Close |