Determination of aluminium, calcium, barium in inoculants by inductively coupled plasma atomic emission spectrometry
Received:March 30, 2015  Revised:April 07, 2015
View Full Text  View/Add Comment  Download reader
DOI:
KeyWord:Inductively Coupled Plasma-Atomic Emission Spectrometry(ICP-AES); Inoculants; Aluminium; Calcium; Barium
        
AuthorInstitution
Yang Hai-jiao 宁夏共享集团股份有限公司(宁夏先进铸造技术重点实验室)
Zhao Yong-li 宁夏共享集团股份有限公司检测中心
Li Jin-bao 宁夏共享集团股份有限公司检测中心
Hits: 975
Download times: 223
Abstract:
      The sample was dissolved with nitric acid, hydrofluoric acid and Hydrochloric acid, and processed with perchloric acid to remove silicon and fluorine, then the salts dissolved by hydrochloric acid. Determination of aluminium, calcium, barium in inoculants by the method of inductively coupled plasma atomic emission spectrometry(ICP-AES) was proposed. The content of the elements could be determined correctly and quickly. The recoveries obtained were in the range of 98% to 103%,the precision of the method and the contrast of analytical results were satisfacted.