Determination of Trace Palladium in Chemical Bulk Drug by ICP-MS with Microwave Digestion
Received:October 09, 2015  Revised:December 07, 2015
View Full Text  View/Add Comment  Download reader
DOI:doi:10.3969/j.issn.2095-1035.2016.01.002
KeyWord:microwave digestion; ICP-MS; Chemical Bulk Drug; Pd
     
AuthorInstitution
LI Li 青海出入境检验检疫局 综合技术中心,西宁
ZHAO Guili 青海出入境检验检疫局 综合技术中心,西宁
Hits: 1434
Download times: 0
Abstract:
      A method for the determination of Pd in Hydrocodone Bitartrate bulk drug by ICP-MS was established in this work.The samples were pretreated by microwave-assisted digestion method and the contents of Pd were determined by ICP-MS under the optimized conditions. The detection limit was 0. 002 9 μg/L with 94.90% ~ 103.4% recovery. The relative standard deviation (RSD) was less than 3%. This method is simple, rapid, sensitive and accurate for the determination of trace Pd in Hydrocodone Bitartrate bulk drugs, and it is also promising in simultaneously determining the contents of other heavy metals in similar chemical bulk drugs.