Determination of Trace Elements in Zinc Ingots by Inductively Coupled Plasma Atomic Emission Spectrometry(ICP-AES)
Received:May 20, 2016  Revised:October 13, 2016
View Full Text  View/Add Comment  Download reader
DOI:doi:10.3969/j.issn.2095-1035.2017.01.005
KeyWord:inductively coupled plasma atomic emission spectrometry; zinc ingot; trace
        
AuthorInstitution
ZANG Zhenjuan 国家铜铅锌及制品质量监督检验中心,安徽 铜陵
JIN Xiaofeng 中国冶金地质总局 山东局测试中心,山东省地质分析测试工程实验室,济南
ZHOU Leiling 中国冶金地质总局 山东局测试中心,山东省地质分析测试工程实验室,济南
Hits: 1053
Download times: 0
Abstract:
      In this article, we determined the concentrations of lead(Pb), cadmium(Cd), iron(Fe), copper(Cu) and tin(Sn) in zinc ingots by inductively coupled plasma atomic emission spectrometry(ICP-AES). The samples were dissolved by nitric acid(1+1), and then the above-mentioned elements were measured in diluted nitric acid medium. The relative standard deviations(RSDs, n=3) were all less than 1.7%. The determination results were consistent with those of the certified reference materials of zinc (BYG0505). With this proposed method, multiple impurity elements in zinc ingots can be simultaneously analyzed with good accuracy and fast speed.