Determination of CO2 in Silicate by Carbon Sulfur Photoelectric Analyzer-Non Aqueous Titration Method
Received:May 17, 2016  Revised:July 07, 2016
View Full Text  View/Add Comment  Download reader
DOI:doi:10.3969/j.issn.2095-1035.2017.01.017
KeyWord:carbon sulfur photoelectric analyzer; non-aqueous titration; CO2
        
AuthorInstitution
ZHANG Yan 1 陕西省矿产资源勘查与综合利用重点实验室 西安 ;2 陕西省地质矿产实验研究所 西安
HAO Hui 1 陕西省矿产资源勘查与综合利用重点实验室 西安 ;2 陕西省地质矿产实验研究所 西安
LIU Xiaoyan 1 陕西省矿产资源勘查与综合利用重点实验室 西安 ;2 陕西省地质矿产实验研究所 西安
Hits: 1099
Download times: 0
Abstract:
      A carbon and sulfur photoelectronic analyzer was applied for the determination of CO2 in silicate samples, which led to a clearer titration end point. The relative error and the relative standard deviation (RSD) for the measurements of the national certified reference material were 0.86%-7.4% and 2.3%-7.9%, respectively. The method can meet the requirements of the accuracy and precision, which are defined in the quality management standard for geology mineral resources laboratory tests (DZ/T0130-2006) .