Determination of Thallium in Lead-Zinc Concentrate by Inductively Coupled Plasma Atomic Emission Spectrometry |
Received:August 21, 2016 Revised:October 09, 2016 |
View Full Text View/Add Comment Download reader |
DOI:doi:10.3969/j.issn.2095-1035.2017.02.003 |
KeyWord:lead-zinc concentrate; thallium; ICP-AES |
Author | Institution |
LI Yanqun |
株洲冶炼集团股份有限公司,湖南 株洲 |
XIANG Delei |
株洲冶炼集团股份有限公司,湖南 株洲 |
|
Hits: 1064 |
Download times: 0 |
Abstract: |
The contents of Tl in Lead-Zinc concentrate samples were determined by ICP-AES after the samples were dissolved in nitric acid, hydrochloric acid, hydrofluoric acid and perchloric acid and the solution was diluted. The measurement range was from 0.005% to 0.05%. By experiments, the appropriate solution acidity and the correction method for interference elements were determined. The relative standard deviations(n=7) were between 1.3% and 4.8%, and the recoveries were in the range from 93.7% to 102%. ?Due to less interference, fast speed and accuracy, the method is applicable for the determination of Tl concentration for the batch samples of Lead-Zinc concentrate. |