Determination of 19 Elements in Fish Feed by Wet Digestion-ICP-AES
Received:July 15, 2016  Revised:October 21, 2016
View Full Text  View/Add Comment  Download reader
DOI:doi:10.3969/j.issn.2095-1035.2017.02.018
KeyWord:ICP-AES; fish feed; multi-element determination; wet-digestion
        
AuthorInstitution
LUO You 中南大学 化学化工学院, 长沙
DENG Feiyue 中南大学 化学化工学院, 长沙
MENG Shixian 贵州中烟工业有限责任公司 贵州 遵义
Hits: 999
Download times: 0
Abstract:
      A new method for simultaneous determination of 19 elements (Cr, Ni, Cd, Pb, and As, ect.) in fish feed was established by ICP-AES together with sample wet digestion method. Under optimized conditions, the correlation coefficients of the calibration curves for all analyzed elements were over 0.999, the standard addition recovery was in the range of 90.74% to 107.0%, and the relative standard deviation was between 0.48% and 9.7%. The national standard material GBW07602(GSV-1) was used to validate the method. The result was consistent with the standard value.The method is simple, quick, accurate, reliable, and suitable for the routine analysis of fish feed.