Pressed Powder Method for the determination of CaF2, CaCO3, S, Fe and SiO2 in Fluorite by XRF
Received:February 21, 2017   Revised:April 07, 2017         Published Online:September 30, 2017
View Full Text  View/Add Comment  Download reader
DOI:doi:10.3969/j.issn.2095-1035.2017.03.014
KeyWord:fluorite; fluorine; pressed powder method; XRF
           
AuthorInstitution
BU Zhaojie Lanzhou LS Test Technology Co. Ltd., Lanzhou, Gansu ,China
WANG Xiaoxuan Lanzhou LS Test Technology Co. Ltd., Lanzhou, Gansu ,China
HUANG Jianqiang Lanzhou LS Test Technology Co. Ltd., Lanzhou, Gansu ,China
TAN Zhaojun Lanzhou LS Test Technology Co. Ltd., Lanzhou, Gansu ,China
Hits: 2234
Download times: 0
Abstract:
      X-ray fluorescence (XRF) spectrometry is a rapid spectroscopic analysis method, although its precision was influenced by the matrix effect and inhomogeneity of elements in samples. The samples was prepared by the grinding the samples and then pressing the power. Content of F, Ca, S, Fe and SiO2 in fluorites were rapidly determined by XRFS. Furthermore, CaF2 and CaCO3 content was also obtained. The results of precision tests indicated that the relative standard deviation (RSD, n=10) of elements to be tested were lower than 0.66%, which meets the testing requirements for these species in fluorite.