Determination of Potassium Oxide Content in Water Soluble Fertilizer by Flow Analytical System |
Received:October 11, 2017 Revised:October 29, 2017 |
View Full Text View/Add Comment Download reader |
DOI:doi:10.3969/j.issn.2095-1035.2018.02.015 |
KeyWord:Continuous Flow Analyzer;Water Soluble Fertilizer;Potassium Oxide |
Author | Institution |
SUN Lili |
云南省化工研究院 |
ZHANG Zhijie |
云南云天化股份有限公司 |
GUI Suping |
云南省化工研究院 |
MAO Hongxiang |
云南省化工研究院 |
|
Hits: 961 |
Download times: 729 |
Abstract: |
The potassium oxidecontent in water soluble fertilizer is determined by AA3 type continuous flow analyzer and compared with NY/T 1977-2010 method. The results show that the correlation coefficient of standard curve is 0.9995 between 0 to 400 mg /L.Compared with weight method, the results is close, below 0.23%.The RSD of ten times repeated measurement was 0.22%. the adding standard recovery rate is 98.2% - 100.6%,there is no significant difference in both methods. |
Close |