Preparation SandSpropertySofElectrographic gel style All-State-Ag/AgCl Reference Electrode
Received:December 26, 2017  Revised:December 26, 2017
View Full Text  View/Add Comment  Download reader
DOI:doi:10.3969/j.issn.2095-1035.2018.03.016
KeyWord:Electrographic gel; impermeable; All-state-reference electrode; Ag/AgCl
              
AuthorInstitution
Yin Nana 辽宁师范大学化学化工学院 辽宁大连
LI Mingqian 辽宁师范大学化学化工学院 辽宁大连
DAI Xin 辽宁师范大学化学化工学院 辽宁大连
LI Yuying 辽宁师范大学化学化工学院 辽宁大连
HE Chunxiang 辽宁师范大学化学化工学院 辽宁大连
Hits: 2046
Download times: 753
Abstract:
      The praparation method and property of a electrographic gel style all-state-Ag/AgCl reference electrode(ASRE) were studied, and the electrographic gel was consisting of AgCl-KCl-polyvinyl alcohol-agar. The ASRE cooperatedSwith pH glass electrode was used for the determination of pH of phosphate buffered solution in the range of pH 2.00~12.00, the results were compared with saturated calomel electrode(SCE) as RE, relative errors(Er) were -0.8% ~ 0.8%. The ASRE cooperatedSwith chloride ion selective electrode was used for the determination of Cl- in the range of concentretions 1.000×10-1~1.000×10-4 mol/L, the electromotive force(△E) has a linear relationship with -lgc(Cl-) with the slope of -49.3, coincidental with -49.9 when SCE as RE. The ASRE can straighway be used for determination of Cl- withoutSdouble salt bridge. In the range of temperature 10~80℃, △E were linear with T, and the temperature coefficient was 0.123 mV/℃. It is expected that the ASRE can be used for off-line or on-line detection.
Close