A Multi - Calibration Curve - Full Spectrum ac/dc Arc Emission Spectrometry was Used to Determine the Plumbum Content in Geological Samples.
Received:January 26, 2018  Revised:January 30, 2018
View Full Text  View/Add Comment  Download reader
DOI:doi:10.3969/j.issn.2095-1035.2018.03.005
KeyWord:multiple calibration curve method;dynamic range;conversion value
                    
AuthorInstitution
WU Dongmei 北京北分瑞利分析仪器(集团)有限责任公司
ZHAO Yanqiu 北京北分瑞利分析仪器(集团)有限责任公司
FU Guoyu 北京北分瑞利分析仪器(集团)有限责任公司
YU Shushi 北京北分瑞利分析仪器(集团)有限责任公司
YU Zhongqi 北京北分瑞利分析仪器(集团)有限责任公司
TAN Ting 北京北分瑞利分析仪器(集团)有限责任公司
LIU Qianyang 北京北分瑞利分析仪器(集团)有限责任公司
Hits: 1286
Download times: 708
Abstract:
      Instrument analysis, when the element content changes greatly, part of the low content and high content of sample is hard to get ideal results. Through the use of multiple calibration curve to enlarge the dynamic range of element measurement. On the AES-8000 full spectrum ac/dc arc emission spectrometer, using this method to test the different content of plumbum elements in the geological samples, and the results show that the measured Plumbum content is closer to the recommended value, i.e. the accuracy is better. In order to broaden the dynamic range, the sample is measured by the first curve of high sensitivity, if the concentration is greater than the conversion value, measured by the second curve of low sensitivity. Cohesion between two curves should be fully considered, the conversion value should be in the linear part of the two curves. Using multiple calibration curve method, setting reasonable conversion value, the high and low content of samples of ideal test results can be obtained at the same time.