Determination of Lanthanum , Cerium in Copper–steel clad plates for bearings by Inductively coupled plasma atomic emission spectroscopy |
Received:July 05, 2018 Revised:July 12, 2018 |
View Full Text View/Add Comment Download reader |
DOI:10.3969/j.issn.2095-1035.2018.06.013 |
KeyWord:inductively coupled plasma atomic emission spectroscopy; Copper–steel clad plates for bearings; La and Ce; |
Author | Institution |
NIE Xiaoming |
广西冶金产品质量检验站 |
HE Xiaohu |
广西冶金产品质量检验站 |
HU Yongmei |
广西冶金产品质量检验站 |
WEI Li |
广西冶金产品质量检验站 |
|
Hits: 1076 |
Download times: 767 |
Abstract: |
Copper–steel clad plates for bearings was digested in nitric acid and hydrochloric acid, with Rh as internal standard, using inductively coupled plasma atomic emission spectroscopy (ICP-AES) for determining the intensity of characteristic line of elements to be measured in specimen solution.The contents of La and Ce in specimen were calculated by standard curve method.The precisions of 1 # and 2# copper–steel clad plates for bearings ( RSD ) were in the rate of 0.98% ~ 1.08%,and the recoveries rate were between 97.0% and 104.6%.So the method is accurate and reliable. |