Determination of Tin in Tin Ores by Inductively Coupled Plasma-Atomic Emission Spectrometry with Alkali Fusion
Received:July 10, 2018  Revised:July 18, 2018
View Full Text  View/Add Comment  Download reader
DOI:doi:10.3969/j.issn.2095-1035.2018.05.008
KeyWord:Tin Ore; Inductively Coupled Plasma-Atomic Emission Spectrometry
           
AuthorInstitution
Xiao Liu-jing 湖南省地质测试研究院
Tang Xing 湖南省地质测试研究院
Wu Yu-hua 湖南省地质测试研究院
Dai Chang-wen 湖南省地质测试研究院
Hits: 1101
Download times: 673
Abstract:
      A method of determination tin in tin ores was established by inductively coupled plasma-atomic emission spectrometry. 2g Na2O2 was used to digestion 0.2g samples without another incandesce. The samples were advised to test as soon as possible to avoid hydrolysis. The detection limit of this method is 38.4 mg/kg and the precision of measurement (RSD) less than 5%. Satisfactory results also obtained by testing the certified reference materials.
Close