Determination of sulfur in uranium trioxide separated by D001 cation exchange resin and measured by Inductively coupled plasma emission spectrometry(ICP-OES)Jiang Haiying[1,] Wang Yongzhu2
Received:August 08, 2018  Revised:August 08, 2018
View Full Text  View/Add Comment  Download reader
DOI:
KeyWord:Inductively coupled plasma emission spectrometry; D001 Cation exchange resin ; UO3; sulfur
     
AuthorInstitution
jianghaiying 北京国原新技术有限公司
王永柱 中国劳动关系学院
Hits: 1143
Download times: 680
Abstract:
      A method for the determination of sulfur in uranium trioxide (UO3) was established. The negative and positive ions in UO3 samples were separated by D001 cation exchange resin, The subboiling water was used to balanced chromatographic column and rinsed the solution .The content of sulfur elements was determined by ICP-OES. Under the experimental conditions, the detection limit of sulfur element is 0.24 μg/mL, the recovery rate is between 91.47% ~ 98.49%, and the relative standard deviations of detection results was 3.07% and 3.15% respectively.