Determination of Arsenic in high purity calcium aluminate powder by hydride formation-atomic fluorescence spectrometry
Received:September 06, 2018  Revised:December 04, 2018
View Full Text  View/Add Comment  Download reader
DOI:
KeyWord:atomic fluorescence spectrometry; calcium aluminate; arsenic
  
AuthorInstitution
XUE Ning 中国铝业郑州有色金属研究院有限公司 河南郑州
Hits: 1034
Download times: 692
Abstract:
      The determination of arsenic in high purity calcium aluminate powder was studied by hydride generation-atomic fluorescence spectrometry. The sample was dissolved in hydrochloric acid(1 +1) and nitric acid(1 +1), the influence of acid medium and acidity, the amount of thiourea and L-ascorbic acid were studied. Under the optimized experimental conditions, the linear correlation coefficient of arsenic standard curve is greater than 0.9995. Precision and recovery tests were carried out on the samples, the relative standard deviations (RSD,n=6) were between 1.84% and 2.55%,the standard recovery rate were between 98.5% and 107.5% This method could satisfy with the determination of arsenic in high calcium aluminate powder.