Determination of arsenic in gold concentrate of Russia by ICP-OES |
Received:October 16, 2018 Revised:December 04, 2018 |
View Full Text View/Add Comment Download reader |
DOI: |
KeyWord:Inductively coupled plasma emission spectra (ICP-OES); recovery rate; Arsenic |
Author | Institution |
Luo Hai Xia |
北矿检测技术有限公司金属矿产资源评价与分析检测北京市重点实验室 |
|
Hits: 1246 |
Download times: 764 |
Abstract: |
The gold concentrate samples were dissolved with HCl-HNO3 mixed acid and determined by ICP-OES. Under the selected operating conditions, the detection limit of arsenic is 0.018μg/mL. The relative standard deviation (n =11) was less than 2% RSD and the recovery was between 98.2 % and 103.0%. Using this method to determine arsenic content in gold concentrate is easy to operate and can meet daily production needs. |
Close |