Determination of Lanthanum, Cerium, Yttrium in Nodulizer by Inductively Coupled Plasma Atomic Emission Spectrometry
Received:January 29, 2019  Revised:January 31, 2019
View Full Text  View/Add Comment  Download reader
DOI:doi:10.3969/j.issn.2095-1035.2019.03.014
KeyWord:Inductively Coupled Plasma-Atomic Emission Spectrometry(ICP-AES);Nodulizer;Rare Earth;Lanthanum
        
AuthorInstitution
Yang Hai-jiao 共享装备股份有限公司检测中心
Luo Kun 共享装备股份有限公司检测中心
Zhao Yong-li 共享装备股份有限公司检测中心
Hits: 1530
Download times: 801
Abstract:
      The sample was dissolved with nitric acid,hydrofluoric acid and hydrochloric acid,and processed with perchloric acid to remove silicon and fluorine,then the salts dissolved by hydrochloric acid.Determination of lanthanum,cerium,yttrium in nodulizer by the method of inductively coupled plasma atomic emission spectrometry(ICP-AES) was proposed.The content of the elements could be determined correctly and quickly.The recoveries obtained were in the range of 97% to 100%,the precision of the method was satisfacted.