Preparation and Electrochemical study of N-N Dimethylformamide / Nano-silver modified electrode |
Received:March 08, 2019 Revised:April 11, 2019 |
View Full Text View/Add Comment Download reader |
DOI:10.3969/j.issn.2095-1035.2019.04.013 |
KeyWord:n-n dimethylformamide; Nano silver; Glass electrode (ITO);Electrochemical response; Metformin hydrochloride |
Author | Institution |
chenzhangwu |
大连大学 环境与化学工程学院 |
zhouhuiyuan |
大连诚泽检测有限公司 |
xiongkaipeng |
大连大学 环境与化学工程学院 |
shenguijun |
大连大学 环境与化学工程学院 |
|
Hits: 1910 |
Download times: 1607 |
Abstract: |
N-N dimethylamide/nano-silver modified electrode was used to modify conductive glass (ITO) with N-N dimethylformamide and nano-silver. The electrochemical behavior of the electrode was studied. In PBS with pH 6.64 as buffer solution, metformin hydrochloride showed a good linear relationship with oxidation peak current within the concentration range of 0.075~0.2 mg/mL while linear equations was y = 317.19x + 30.07 6, and the minimum detection limit was 0.000 7 mg/ mL. The correlation linear coefficient r squared is 0.959 5. Parallel measurement of 5 sets of data with an RSD value of 0.75 per cent. In metformin hydrochloride sustained-release tablet solution, the average recovery rate was 100.96 per cent after 5 times of standard addition. The results of precision and accuracy analysis showed that this method can be applied to the determination of metformin hydrochloride in medicine. |
Close |