Determination of gold content in copper anode mud by inductively coupled plasma emission spectrometry
Received:June 17, 2019  Revised:June 21, 2019
View Full Text  View/Add Comment  Download reader
DOI:doi:10.3969/j.issn.2095-1035.2019.05.011
KeyWord:ICP-AES;copper anode mud;gold
              
AuthorInstitution
LONG-Haizhen 西部矿业科技发展有限公司
KONG-Huimin 西部矿业科技发展有限公司
YUE-Ping 西部矿业科技发展有限公司
WANG-Jingfeng 西部矿业科技发展有限公司
HUANG-Hong 西部矿业科技发展有限公司
Hits: 974
Download times: 865
Abstract:
      The sample of copper anode mud from a copper smelter was used to test the gold content.Dissolve the sample with aqua regia,set the capacity to 100mL capacity bottle under 3%(V/V) aqua regia acidity, determined by inductively coupled plasma emission spectrometry. After experimental research, 242.795nm was selected as the analytical spectral line. The detection limit of the method was(N=11) 0.09μg/mL, the lower limit was determined to be 0.27μg/mL, and the standard recovery rate was between 97% and 107%. The linear correlation coefficient is: 0.9999928. Analysis results meet requirements.