Determination of gold content in copper anode mud by inductively coupled plasma emission spectrometry |
Received:June 17, 2019 Revised:June 21, 2019 |
View Full Text View/Add Comment Download reader |
DOI:doi:10.3969/j.issn.2095-1035.2019.05.011 |
KeyWord:ICP-AES;copper anode mud;gold |
Author | Institution |
LONG-Haizhen |
西部矿业科技发展有限公司 |
KONG-Huimin |
西部矿业科技发展有限公司 |
YUE-Ping |
西部矿业科技发展有限公司 |
WANG-Jingfeng |
西部矿业科技发展有限公司 |
HUANG-Hong |
西部矿业科技发展有限公司 |
|
Hits: 974 |
Download times: 865 |
Abstract: |
The sample of copper anode mud from a copper smelter was used to test the gold content.Dissolve the sample with aqua regia,set the capacity to 100mL capacity bottle under 3%(V/V) aqua regia acidity, determined by inductively coupled plasma emission spectrometry. After experimental research, 242.795nm was selected as the analytical spectral line. The detection limit of the method was(N=11) 0.09μg/mL, the lower limit was determined to be 0.27μg/mL, and the standard recovery rate was between 97% and 107%. The linear correlation coefficient is: 0.9999928. Analysis results meet requirements. |
Close |