Determination of 12 elements in Nickel-based Alloys by spark discharge atomic emission spectrometry
Received:April 01, 2020  Revised:April 13, 2020
View Full Text  View/Add Comment  Download reader
DOI:10.3969/j.issn.2095-1035.2021.01.011
KeyWord:Nickel-base alloy; Spark discharge atomic emission spectrometry; Element; Determination
              
AuthorInstitution
fengxiumei 江阴市产品质量监督检验所
CHEN Lianfang 国家船舶材料质量监督检验中心江苏 江苏江阴
CHEN Jun (1. 江阴市产品质量监督检验所,江苏 江阴 ;2.国家船舶材料质量监督检验中心江苏,
changshouqin 国家船舶材料质量监督检验中心江苏 江苏江阴
ganmeilu 国家船舶材料质量监督检验中心江苏 江苏江阴
Hits: 717
Download times: 817
Abstract:
      The contents of chemical elements and their changes affect the properties of nickel-based alloy materials. Therefore, the accurate detection and strict control of the chemical composition of nickel-based alloys are of great significance to improve and guarantee the properties of nickel-based alloys. By optimizing the parameters of the instrument and using sample control method, a method for determining the contents of Al, C, Cr, Co, Cu, Fe, Nb, P, Si, S, W and Mo elements in nickel base alloy by spark discharge atomic emission spectrometry was established. The optimized test conditions are as follows: (1) argon-blowing time: 6 s; (2) pre-combustion time: 6 s; The integral time: 8 s. Three kinds of conventional nickel base alloy samples were analyzed by this method. The measured values are consistent with the results of the current national standard analytical methods.