Inductively Coupled Plasma Atomic Emission SpectroscopyDetermination of multiple elements in Water-chilled dreg with Microwave Digestion
Received:June 14, 2020  Revised:June 24, 2020
View Full Text  View/Add Comment  Download reader
DOI:doi:10.3969/j.issn.2095-1035.2020.05.0011
KeyWord:Inductively;coupled plasma;atomic emission;spectroscopy, microwave;digestion, Water-chilled;dreg。
     
AuthorInstitution
zuohongyi 深圳市中金岭南有色金属股份有限公司韶关冶炼厂
shishilong 深圳市中金岭南有色金属股份有限公司韶关冶炼厂
Hits: 1144
Download times: 662
Abstract:
      The sample was digested by microwave with Aqua and hydrofluoric acid, and multiple elements in Water-chilled dreg was determined by Inductively coupled plasma atomic emission spectroscopy. The sample can be dissolved completely in an airtight container at 185 °C for 60 Min, the interference of the composition of the digestion solution can be eliminated by selecting the appropriate spectrum line, the recovery rate is 95.72%~106.89%, the relative standard deviation is 0.68%~8.65% (N = 5) , the detection limit meet the analytical requirements。 The results show that the method is fast,simple and accurate,and the method can be used for the determination of multiple elements in Water-chilled dreg.