Determination of cobalt, copper and manganese in cobalt concentrate by wavelength dispersive X-ray fluorescence spectrometry
Received:January 15, 2021  Revised:June 04, 2021
View Full Text  View/Add Comment  Download reader
DOI:10.3969/j.issn.2095-1035.2021.04.003
KeyWord:x-ray fluorescence spectrometry; cobalt concentrate; cobalt; copper; manganese; fusion sample preparation
                       
AuthorInstitution
FENG YA HUI 南京海关工业产品检测中心
HOU JIAN JUN 南京海关工业产品检测中心
PAN Shenglin 南京海关工业产品检测中心
DAI Dongqing 南京海关工业产品检测中心
ZHANG XIU 南京海关工业产品检测中心
WANG Xuning 南京金利检验有限公司,
XU REN FU 南京海关工业产品检测中心
MENG QIN SEN 南京金利检验有限公司
Hits: 104
Download times: 78
Abstract:
      A method for the determination of cobalt, copper and manganese in cobalt concentrate by wavelength dispersive x-ray fluorescence spectrometry was established. The artificial standard sample was prepared by melting high-purity oxides with lithium tetraborate and lithium metaborate. The calibration curve was made with ytterbium oxide as internal standard. The results show that the calibration curve of each element has a wide linear range, the correlation coefficient is greater than 0.999, and the detection limits of cobalt, copper and manganese are 0.002%, 0.001% and 0.001% respectively. The test accuracy and precision are reliable, and the method meets the needs of the industry.