Determination of Trace Pb and Cd in High Salinity Water by ICP-AES with Sulfhydryl Cotton Separation and Preconcentration
Received:May 28, 2013  Revised:June 21, 2013
View Full Text  View/Add Comment  Download reader
DOI:doi:10.3969/j.issn.2095-1035.2013.04.007
KeyWord:sulfhydryl cotton; lead; cadmium; inductively coupled plasma atomic emission spectrometry; high salinity
           
AuthorInstitution
ZHANG Ning 河北省地矿中心实验室,河北 保定
GUO Xiuping 河北省地矿中心实验室,河北 保定
SHEN Yumin 河北省地矿中心实验室,河北 保定
JIANG Yunjun 河北省地矿中心实验室,河北 保定
Hits: 2315
Download times: 271
Abstract:
      In this paper, trace Pb and Cd in high salinity water was determined by by ICP-AES. Sulfhydryl cotton was used for sample separation and preconcentration purpose. The results show that at pH=7 the best enrichment effect for both trace Pb and Cd was achieved and the matrix elements were also separated after the colum was washed with 1.5 mol/L HCl solution. The recoveries for Pb and Cd were in the range of 95.0%~105.0% and the RSDs were between 3.8% to 9.7%.