Determination of Aluminum, Copper, Lead, Iron, Cadmium, Manganese in Zinc Oxide Produced by Direct Process by Inductively Coupled Plasma Optical Emission Spectrometry(ICP-OES)
Received:June 14, 2013   Revised:July 10, 2013         Published Online:December 19, 2013
View Full Text  View/Add Comment  Download reader
DOI:doi:10.3969/j.issn.2095-1035.2013.04.011
KeyWord:ICP-OES; zinc oxide produced by direct process; aluminum, copper, lead, iron, cadmium, manganese; determination
        
AuthorInstitution
ZHANG Shitao Yantai Penghui Copper Industry Co., Ltd., Yantai, Shandong
ZHANG Lianxiang Beijing General Research Institute of Mining and Metallurgy, Beijing
FENG Xianjin Beijing General Research Institute of Mining and Metallurgy, Beijing
Hits: 3602
Download times: 1244
Abstract:
      A method for the determination of Al, Cu, Pb, Fe, Cd, Mn in zinc oxide produced by direct process by Inductively Coupled Plasma Optical Emission Spectrometry(ICP-OES) was established. Sample dissolution method was determined and analytical spectral lines were selected. Method precision and accuracy were tested. The results show that the relative standard deviation of each element is between 1.10% and 7.63% and standard addition recovery is between 92% and 116%. The determination results are consistent with those by other traditional analysis methods. The established method is accurate, fast and suitable for simultaneous determination of multi-element in zinc oxide produced by direct process.