Determination of Sulfur in Cobalt Carbonate by Infrared Absorption Spectrometry Method after High Frequency Combustion
Received:May 13, 2013   Revised:September 06, 2013         Published Online:December 19, 2013
View Full Text  View/Add Comment  Download reader
DOI:doi:10.3969/j.issn.2095-1035.2013.04.012
KeyWord:high frequency combustion;cobalt carbonate;fluxing medium;sulfur
  
AuthorInstitution
LUO Yueying Testing Center of Yingde Jiana Metal Technology Co.,Ltd.,Yingde,Guangdong ,China
Hits: 3276
Download times: 1167
Abstract:
      Sulfur in cobalt carbonate samples was determined by infrared absorption spectrometry method with high frequency combustion. In the process of determination, the samples were firstly burned and then melted in pure iron and tungsten fluxing medium. This method solved the problems of moisture interference and incomplete sample melting. The method can accurately determine the sulfur with mass fraction in cobalt carbonate as 0.001%~0.01%.The spiked recovery rates are between 98.4%~101.4% and relative standard deviations are between 2.8%~6.3%.