Determination of Heavy Metal Lead in Neodymium-Iron-Boron with ICP-OES
Received:November 14, 2018  Revised:March 07, 2019
View Full Text  View/Add Comment  Download reader
DOI:doi:10.3969/j.issn.2095-1035.2019.02.003
KeyWord:ICP-OES;neodymium-iron-boron; heavy metal;lead
  
AuthorInstitution
zhuxiaolong 宁波市计量测试研究院
Hits: 1086
Download times: 689
Abstract:
      The sample of neodymium-iron-boron magnet was digested by nitric acid. An analysis method was established for determination of heavy metal lead in neodymium-iron-boron magnet with inductively coupled plasma optical emission spectrometer(ICP-OES). The working conditions of the instrument and analysis wavelength were discussed. Results showed that the linear correlation coefficient was 0.99993, the detection limit was 0.009 μg/mL, the recovery of standard addition was 104 %, the accuracy and precision were all less than 3 %. The detection limit of the method was low and met the actual analysis requirements. This method could quickly, conveniently and accurately detect the heavy metal lead in neodymium-iron-boron magnet.