Determination of the 6 Impurity metal ions in handmade paper by ICP-MS
Received:November 16, 2018   Revised:March 20, 2019   Accepted:December 07, 2018      Published Online:May 27, 2019
View Full Text  View/Add Comment  Download reader
DOI:doi:10.3969/j.issn.2095-1035.2019.02.001
KeyWord:ICP-MS; Traditional handmade paper; Impurity metal
                 
AuthorInstitution
luchq Institute of Quality Inspection of Light Industry
li wen-hui Institute of Quality Inspection of Light Industry
zhang ruo-hong Institute of Quality Inspection of Light Industry
sun kan Institute of Quality Inspection of Light Industry
luo chan Institute of Quality Inspection of Light Industry
wan feng Institute of Quality Inspection of Light Industry
Hits: 3210
Download times: 1677
Abstract:
      Samples of handmade paper was pretreated by microwave assisted digestion, and the contents of As, Pb, Cr, Ni, Sb and Cd in the handmade paper were determined by ICP-MS. Kinetic energy discrimination (KED) were used to overcome disturbance of ions,such as ArCl. The parameters of the test methods were optimized seriously, including RPq and affects of different Neb gas flow rate. Detection limits of the 6 elements were in the ranges of 0.003 mg.kg-1~0.02 mg.kg-1, 3.0mg.kg-1 and 7.0 mg.kg-1 mix standard solution were added in the samples respectively. Values of recovery and RSD’s were found in the range of 88.9%~107.3% and 1.7%~2.6%. As, Pb, Cr, Ni and Sb, were detected in some of the samples. The method is fast, accurate and has a low detection limit. It is suitable for determination of impurity metal ions in handmade paper